                              INSTRUMENT CONTROL PARAMETERS:    MSD5975
                              -----------------------------------------

   C:\MSDCHEM\1\METHODS\SCREENING.M
      Mon Dec 03 13:20:37 2012

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

Oven
Equilibration Time                           0.5 min
Max Temperature                              325 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    50 C for 0.5 min
    then 75 C/min to 100 C for 0.8 min
    then 15 C/min to 300 C for 20 min
Run Time                                     35.3 min
2 min (Post Run)                             320 C




Front Injector
Syringe Size                                 10 L
Injection Volume                             1 L
Solvent A Washes (PreInj)                    5
Solvent A Washes (PostInj)                   5
Solvent A Volume                             8 L
Solvent B Washes (PreInj)                    5
Solvent B Washes (PostInj)                   5
Solvent B Volume                             8 L
Sample Washes                                0
Sample Wash Volume                           8 L
Sample Pumps                                 0
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 L/min
Solvent Wash Dispense Speed                  6000 L/min
Sample Wash Draw Speed                       300 L/min
Sample Wash Dispense Speed                   6000 L/min
Injection Dispense Speed                     6000 L/min
Viscosity Delay                              0 sec
Sample Depth                                 Disabled
Injection Type                               Standard
L1 Airgap                                    0 L


Sample Overlap
Sample overlap is not enabled



Front SSZ Inlet He
Mode                                         Pulsed Splitless
Heater                                       On    300 C
Pressure                                     On    9.7853 psi
Total Flow                                   On    103.3 mL/min
Septum Purge Flow                            On    2.1035 mL/min
Gas Saver                                    On    20 mL/min After 2 min
Injection Pulse Pressure                     25 psi Until 0.5 min
Purge Flow to Split Vent                     100 mL/min at 1 min


Thermal Aux 1 (MSD Transfer Line)
Heater                                       On
Temperature Program                          On
    250 C for 0 min
Run Time                                     35.3 min


Column #1
Agilent 19091S-433: 1839.63508
HP-5MS  5% Phenyl Methyl Silox
325 C: 30 m x 250 m x 0.25 m
In: Front SSZ Inlet He
Out: Vacuum

(Initial)                                    50 C
Pressure                                     9.7853 psi
Flow                                         1.2 mL/min
Average Velocity                             39.923 cm/sec
Holdup Time                                  1.2524 min
Flow Program                                 On
    1.2 mL/min for 0 min
Run Time                                     35.3 min
2 min (Post Run)                             1.5 mL/min



Signals
Signal #1: None
Signal #2: None
Signal #3: None
Signal #4: None

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 4.10 min

EMV Mode                 : Gain Factor
Gain Factor              : 1.00
Resulting EM Voltage     : 1271

[Scan Parameters]

Low Mass                 : 43.0
High Mass                : 650.0
Threshold                : 50
Sample #                 : 2       A/D Samples    4
Plot 2 low mass          : 50.0
Plot 2 high mass         : 550.0

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US11409902
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      26.940
 IONFOCUS    :      90.157
 ENTRANCE_LE :      25.500
 EMVOLTS     :    1129.412
                               Actual EMV  :    1270.59
                               GAIN FACTOR :       0.99
 AMUGAIN     :    2089.000
 AMUOFFSET   :     123.625
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      19.827
 MASSGAIN    :    -554.000   
 MASSOFFSET  :     -35.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                                 END OF INSTRUMENT CONTROL PARAMETERS
                                 ------------------------------------