                        INSTRUMENT CONTROL PARAMETERS:    Instrument #2
                        -----------------------------------------------

   C:\MSDCHEM\1\METHODS\drug.m
      Mon Feb 11 12:26:25 2013

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  Manual
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

Oven
Equilibration Time                           3 min
Max Temperature                              325 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    100 C for 3 min
    then 10 C/min to 280 C for 20 min
Run Time                                     41 min
Cryo                                         Off

Front Injector
Syringe Size                                 10 L
Injection Volume                             1 L
Solvent A Washes (PreInj)                    2
Solvent A Washes (PostInj)                   0
Solvent A Volume                             8 L
Solvent B Washes (PreInj)                    0
Solvent B Washes (PostInj)                   0
Solvent B Volume                             8 L
Sample Washes                                0
Sample Wash Volume                           8 L
Sample Pumps                                 3
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 L/min
Solvent Wash Dispense Speed                  6000 L/min
Sample Wash Draw Speed                       300 L/min
Sample Wash Dispense Speed                   6000 L/min
Injection Dispense Speed                     6000 L/min
Viscosity Delay                              0 sec
Sample Depth                                 Disabled
Injection Type                               Standard
L1 Airgap                                    0.2 L

Sample Overlap
Sample overlap is not enabled

Front SS Inlet He
Mode                                         Splitless
Heater                                       On    280 C
Pressure                                     On    14.058 psi
Total Flow                                   On    19.3 mL/min
Septum Purge Flow                            On    3 mL/min
Gas Saver                                    On    20 mL/min After 2 min
Purge Flow to Split Vent                     15 mL/min at 0.75 min

Thermal Aux 2 (MSD Transfer Line)
Heater                                       On
Temperature Program                          On
    280 C for 0 min
Run Time                                     41 min

Column #1
Agilent 19091J-433: 2080.69177
HP-5  5% Phenyl Methyl Siloxan
325 C: 30 m x 250 m x 0.25 m
In: Front SS Inlet He
Out: Vacuum

(Initial)                                    100 C
Pressure                                     14.058 psi
Flow                                         1.3 mL/min
Average Velocity                             42.521 cm/sec
Holdup Time                                  1.1759 min
Flow Program                                 On
    1.3 mL/min for 0 min
Run Time                                     41 min

Signals
Signal #1: Test Plot                         Save Off
                                             50 Hz

Signal #2: Test Plot                         Save Off
                                             50 Hz

Signal #3: Test Plot                         Save Off
                                             50 Hz

Signal #4: Test Plot                         Save Off
                                             50 Hz

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 3.00 min

EMV Mode                 : Gain Factor
Gain Factor              : 1.00
Resulting EM Voltage     : 1200

[Scan Parameters]

Low Mass                 : 30.0
High Mass                : 600.0
Threshold                : 150
Sample #                 : 2       A/D Samples    4
Plot 2 low mass          : 40.0
Plot 2 high mass         : 550.0

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US12318802
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      27.777
 IONFOCUS    :      90.157
 ENTRANCE_LE :      28.500
 EMVOLTS     :    1035.294
                               Actual EMV  :    1200
                               GAIN FACTOR :       1.06
 AMUGAIN     :    1663.000
 AMUOFFSET   :     120.375
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      17.569
 MASSGAIN    :    -708.000   
 MASSOFFSET  :     -37.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                              END OF INSTRUMENT CONTROL PARAMETERS
                              ------------------------------------