                                       INSTRUMENT CONTROL PARAMETERS:    Instrument #2
                                       -----------------------------------------------

   C:\MSDCHEM\1\METHODS\A_SCR.M
      Thu Aug 22 11:33:10 2013

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

Oven
Equilibration Time                           3 min
Max Temperature                              325 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    100 C for 2 min
#1    then 25 C/min to 250 C for 2 min
#2    then 10 C/min to 280 C for 7 min
Run Time                                     20 min
Cryo                                         Off

Front Injector
Syringe Size                                 10 ?L
Injection Volume                             2 ?L
Solvent A Washes (PreInj)                    1
Solvent A Washes (PostInj)                   5
Solvent A Volume                             8 ?L
Solvent B Washes (PreInj)                    1
Solvent B Washes (PostInj)                   2
Solvent B Volume                             8 ?L
Sample Washes                                0
Sample Wash Volume                           8 ?L
Sample Pumps                                 1
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 ?L/min
Solvent Wash Dispense Speed                  6000 ?L/min
Sample Wash Draw Speed                       300 ?L/min
Sample Wash Dispense Speed                   6000 ?L/min
Injection Dispense Speed                     6000 ?L/min
Viscosity Delay                              0 sec
Sample Depth                                 1 mm
Injection Type                               Standard
L1 Airgap                                    0 ?L

Sample Overlap
Sample overlap is not enabled

Front SS Inlet He
Mode                                         Splitless
Heater                                       On    280 C
Pressure                                     On    10.523 psi
Total Flow                                   On    24 mL/min
Septum Purge Flow                            On    3 mL/min
Gas Saver                                    On    20 mL/min After 2 min
Purge Flow to Split Vent                     20 mL/min at 0.75 min

Thermal Aux 2 (MSD Transfer Line)
Heater                                       On
Temperature Program                          On
    280 C for 0 min
Run Time                                     20 min

Column #1
Agilent 19091J-433: 2080.69177
HP-5  5% Phenyl Methyl Siloxan
325 C: 30 m x 250 ?m x 0.25 ?m
In: Front SS Inlet He
Out: Vacuum

(Initial)                                    100 C
Pressure                                     10.523 psi
Flow                                         1 mL/min
Average Velocity                             37.293 cm/sec
Holdup Time                                  1.3407 min
Flow Program                                 On
    1 mL/min for 0 min
Run Time                                     20 min

Signals
Signal #1: Test Plot                         Save Off
                                             50 Hz

Signal #2: Test Plot                         Save Off
                                             50 Hz

Signal #3: Test Plot                         Save Off
                                             50 Hz

Signal #4: Test Plot                         Save Off
                                             50 Hz

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 3.00 min

EMV Mode                 : Gain Factor
Gain Factor              : 3.00
Resulting EM Voltage     : 1506

[Scan Parameters]

Low Mass                 : 40.0
High Mass                : 550.0
Threshold                : 150
Sample #                 : 2       A/D Samples    4
Plot 2 low mass          : 40.0
Plot 2 high mass         : 550.0

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US12318802
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      34.814
 IONFOCUS    :      90.157
 ENTRANCE_LE :      16.000
 EMVOLTS     :    1505.882
                               Actual EMV  :    1505.88
                               GAIN FACTOR :       2.94
 AMUGAIN     :    1679.000
 AMUOFFSET   :     122.688
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      20.329
 MASSGAIN    :    -725.000   
 MASSOFFSET  :     -37.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                                            END OF INSTRUMENT CONTROL PARAMETERS
                                            ------------------------------------