                        INSTRUMENT CONTROL PARAMETERS:    Instrument #1
                        -----------------------------------------------

   C:\MSDCHEM\1\METHODS\NARCO_MET2.m

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled

=============================================================================
                                6850 GC METHOD
=============================================================================


OVEN
   Initial temp:  100 'C (On)              Maximum temp:  325 'C
   Initial time:  3.00 min                 Equilibration time:  0.00 min
   Ramps:
      #  Rate  Final temp  Final time
      1 25.00      300       20.00
      2   0.0(Off)
   Post temp:  0 'C
   Post time:  0.00 min
   Run time:  31.00 min


INLET (SPLIT/SPLITLESS)                 DETECTOR (NO DET)
   Mode:  Pulsed Splitless                 Temperature:  300 'C (On)
   Initial temp:  280 'C (On)
   Pressure:  21.29 psi (On)
   Pulse pressure:  25.0 psi
   Pulse time:  0.50 min
   Purge flow:  40.0 mL/min
   Purge time:  0.50 min
   Total flow:  43.5 mL/min
   Gas saver:  Off
   Gas type:  Helium


COLUMN                                  SIGNAL
   Capillary Column                        Data rate:  0.1 Hz
   Model Number:  Agilent 19091B-102E      Type:  col comp
   Ultra 2 5% Phenyl Methyl Siloxane       Save Data:  Off
   Max temperature:  325 'C                Zero:  0.0 (Off)
   Nominal length:  25.0 m                 Range:  0
   Nominal diameter:  200.00 um            Fast Peaks:  Off
   Nominal film thickness:  0.33 um        Attenuation:  0
   Mode:  constant flow
   Initial flow:  1.0 mL/min
   Nominal init pressure:  21.30 psi
   Average velocity:  41 cm/sec
   Source:  Inlet
   Outlet:  MSD
   Outlet pressure:  vacuum


COLUMN COMP
   Use:  Valve Box Heater


VALVES                                  POST RUN
                                           Post Time: 0.00 min


TIME TABLE
   Time       Specifier                     Parameter & Setpoint




                               GC Injector


     Front Injector:
        Sample Washes                 0
        Sample Pumps                  2
        Injection Volume           1.00 microliters
        Syringe Size               10.0 microliters
        PreInj Solvent A Washes       2
        PreInj Solvent B Washes       2
        PostInj Solvent A Washes      1
        PostInj Solvent B Washes      1
        Viscosity Delay               0 seconds
        Plunger Speed              Fast
        PreInjection Dwell         0.00 minutes
        PostInjection Dwell        0.00 minutes
        Sampling Depth             -1.0 mm

     Back Injector:
No parameters specified

 Column 1 Inventory Number : 128-8522e
                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : urtune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 4.00 min

EM Absolute              : False
EM Offset                : 235
Resulting EM Voltage     : 1717.6

[Scan Parameters]

Low Mass                 : 40.0
High Mass                : 500.0
Threshold                : 50
Sample #                 : 2       A/D Samples    4

[MSZones]

MS Quad                  : 150 C   maximum 200 C
MS Source                : 230 C   maximum 250 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US30945420
                        -----------------------------

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      34.311
 IONFOCUS    :      86.361
 ENTRANCE_LE :      12.000
 EMVOLTS     :    1482.353
 AMUGAIN     :    2477.000
 AMUOFFSET   :     133.000
 FILAMENT    :       2.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      21.333
 MASSGAIN    :     396.000   
 MASSOFFSET  :     -11.000   


                           END OF TUNE PARAMETERS
                      ------------------------------------



                              END OF INSTRUMENT CONTROL PARAMETERS
                              ------------------------------------