                                        INSTRUMENT CONTROL PARAMETERS:    MS3_Enh
                                        -----------------------------------------

   C:\MSDCHEM\1\DATA\2014-04-23-0838.B\CNBM_HP5MS_DRV.M
      Wed Apr 23 09:24:07 2014

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled

  Sample Prep Method : 2T_100S_MSTFA_50_T80_M5_10.PT
       Last Modified : 2013-11-05-1025 

1. Wash syringe in Back tower, drawing [Method Value]  from
     Wash A vials dispensing into Current waste vial 1 times where
     offset = [Method Value], draw speed = [Method Value],
     dispense speed = [Method Value], viscosity = [Method Value]

2. Move vial from tray vial #131 to back turret position #3

3. Move vial from tray vial #141 to back turret position #2

4. Move vial from front sequence vial to back turret position #1

5. Set the heater to 80 degrees C

6. Dispense 50 ?L from vial Sample 3 to vial Sample 1 on Back tower where
     syringe offset = [Method Value], Draw Speed = [Method Value],
     Dispense Speed = [Method Value], viscosity = [Method Value],
     air gap = [Method Value]

7. Move vial from back turret position #1 to mixer

8. Mix at 1000 rpm 3 times for 20 seconds where
   spin mode = Bi-directional, dwell time = 5 sec

9. Move vial from mixer to heater

10. Heat vial for 150 seconds using current heater temperature

11. Move vial from heater to mixer

12. Mix at 1000 rpm 1 times for 15 seconds where
   spin mode = Bi-directional, dwell time = 5 sec

13. Move vial from mixer to heater

14. Heat vial for 150 seconds using current heater temperature

15. Turn off heater

16. Move vial from heater to mixer

17. Mix at 1000 rpm 3 times for 20 seconds where
   spin mode = Bi-directional, dwell time = 5 sec

18. Wait for 2 minutes

19. Move vial from mixer to back turret position #1

20. Dispense 150 ?L from vial Sample 2 to vial Sample 1 on Back tower where
     syringe offset = [Method Value], Draw Speed = [Method Value],
     Dispense Speed = [Method Value], viscosity = [Method Value],
     air gap = [Method Value]

21. Move vial from back turret position #1 to mixer

22. Mix at 1000 rpm 3 times for 20 seconds where
   spin mode = Bi-directional, dwell time = 5 sec

23. Return all vials

Oven
Equilibration Time                           3 min
Max Temperature                              325 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    65 C for 1 min
    then 20 C/min to 320 C for 11.25 min
Run Time                                     25 min
Cryo                                         Off

Front Injector
Syringe Size                                 10 ?L
Injection Volume                             1 ?L
Solvent A Washes (PreInj)                    0
Solvent A Washes (PostInj)                   5
Solvent A Volume                             8 ?L
Solvent B Washes (PreInj)                    2
Solvent B Washes (PostInj)                   2
Solvent B Volume                             8 ?L
Sample Washes                                0
Sample Wash Volume                           2 ?L
Sample Pumps                                 0
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 ?L/min
Solvent Wash Dispense Speed                  1000 ?L/min
Sample Wash Draw Speed                       300 ?L/min
Sample Wash Dispense Speed                   1000 ?L/min
Injection Dispense Speed                     6000 ?L/min
Viscosity Delay                              3 sec
Sample Depth                                 -2 mm
Injection Type                               2-layer Sandwich
L1 Airgap                                    1 ?L
L2 Volume                                    0.2 ?L
L2 Airgap                                    0.4 ?L

Back Injector
Syringe Size                                 100 ?L
Injection Volume                             10 ?L
Solvent A Washes (PreInj)                    0
Solvent A Washes (PostInj)                   0
Solvent A Volume                             80 ?L
Solvent B Washes (PreInj)                    0
Solvent B Washes (PostInj)                   0
Solvent B Volume                             80 ?L
Sample Washes                                0
Sample Wash Volume                           80 ?L
Sample Pumps                                 0
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      1000 ?L/min
Solvent Wash Dispense Speed                  3000 ?L/min
Sample Wash Draw Speed                       1000 ?L/min
Sample Wash Dispense Speed                   3000 ?L/min
Injection Dispense Speed                     3000 ?L/min
Viscosity Delay                              3 sec
Sample Depth                                 Disabled
Injection Type                               Standard
L1 Airgap                                    0 ?L

Barcode Reader
Barcode heater                               Disabled
Barcode mixer                                Disabled

Sample Overlap
Sample overlap will occur after the previous injection is completed

Front SS Inlet He
Mode                                         Pulsed Splitless
Heater                                       On    250 C
Pressure                                     On    10.068 psi
Total Flow                                   On    54.138 mL/min
Septum Purge Flow                            On    3 mL/min
Septum Purge Flow Mode                       Switched
Gas Saver                                    On    15 mL/min After 4 min
Injection Pulse Pressure                     36.259 psi Until 0.5 min
Purge Flow to Split Vent                     50 mL/min at 1 min

Thermal Aux 2 (MSD Transfer Line)
Heater                                       On
Temperature Program                          On
    290 C for 0 min
Run Time                                     25 min

Column #1
Agilent 19091S-433: 93.92873
HP-5MS  5% Phenyl Methyl Silox
325 C: 30 m x 250 ?m x 0.25 ?m
In: Front SS Inlet He
Out: Vacuum

(Initial)                                    65 C
Pressure                                     10.068 psi
Flow                                         1.1378 mL/min
Average Velocity                             39.157 cm/sec
Holdup Time                                  1.2769 min
Flow Program                                 Off
    1.1378 mL/min for 0 min
Run Time                                     25 min

Signals
Signal #1: None

Signal #2: None

Signal #3: Test Plot                         Save Off
                                             50 Hz

Signal #4: Test Plot                         Save Off
                                             50 Hz

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 3.75 min

EMV Mode                 : Gain Factor
Gain Factor              : 1.50
Resulting EM Voltage     : 2376

[Scan Parameters]

Low Mass                 : 42.0
High Mass                : 700.0
Threshold                : 150
Sample #                 : 2       A/D Samples    4
Plot 2 low mass          : 50.0
Plot 2 high mass         : 650.0

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US71216556
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      34.814
 IONFOCUS    :      90.157
 ENTRANCE_LE :      22.000
 EMVOLTS     :    2341.176
                               Actual EMV  :    2376.47
                               GAIN FACTOR :       1.50
 AMUGAIN     :    1505.000
 AMUOFFSET   :     122.813
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      17.067
 MASSGAIN    :    -919.000   
 MASSOFFSET  :     -40.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                                          END OF INSTRUMENT CONTROL PARAMETERS
                                          ------------------------------------