                              INSTRUMENT CONTROL PARAMETERS:    MSD
                              -------------------------------------

   c:\msdchem\2\METHODS\SCREEN.M
      Thu Sep 11 19:48:28 2014

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

Oven
Equilibration Time                           0.3 min
Max Temperature                              320 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    100 C for 1 min
    then 35 C/min to 300 C for 10 min
Run Time                                     16.714 min




Front Injector
Syringe Size                                 10 ?L
Injection Volume                             1 ?L
Solvent A Washes (PreInj)                    5
Solvent A Washes (PostInj)                   5
Solvent A Volume                             8 ?L
Solvent B Washes (PreInj)                    5
Solvent B Washes (PostInj)                   5
Solvent B Volume                             8 ?L
Sample Washes                                0
Sample Wash Volume                           8 ?L
Sample Pumps                                 2
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 ?L/min
Solvent Wash Dispense Speed                  6000 ?L/min
Sample Wash Draw Speed                       300 ?L/min
Sample Wash Dispense Speed                   6000 ?L/min
Injection Dispense Speed                     6000 ?L/min
Viscosity Delay                              0 sec
Sample Depth                                 Disabled
Injection Type                               Standard
L1 Airgap                                    0.2 ?L


Sample Overlap
Sample overlap is not enabled



Front SSZ Inlet He
Mode                                         Splitless
Heater                                       On    270 C
Pressure                                     On    95 kPa
Total Flow                                   On    23.537 mL/min
Septum Purge Flow                            On    2.2623 mL/min
Gas Saver                                    Off
Purge Flow to Split Vent                     20 mL/min at 1 min


Thermal Aux 1 (MSD Transfer Line)
Heater                                       On
Temperature Program                          On
    280 C for 0 min
Run Time                                     16.714 min


Column #1

320 C: 30 m x 250 ?m x 0.25 ?m
In: Front SSZ Inlet He
Out: Vacuum

(Initial)                                    100 C
Pressure                                     95 kPa
Flow                                         1.2749 mL/min
Average Velocity                             42.108 cm/sec
Holdup Time                                  1.1874 min
Pressure Program                             On
    95 kPa for 0 min
Run Time                                     16.714 min



Signals
Signal #1: Test Plot                         Save Off
                                             50 Hz
Signal #2: Test Plot                         Save Off
                                             50 Hz
Signal #3: Test Plot                         Save Off
                                             50 Hz
Signal #4: Test Plot                         Save Off
                                             50 Hz

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 3.00 min

EMV Mode                 : Gain Factor
Gain Factor              : 1.00
Resulting EM Voltage     : 1541

[Scan Parameters]

Low Mass                 : 41.0
High Mass                : 650.0
Threshold                : 150
Sample #                 : 2       A/D Samples    4

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US11341801
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      29.453
 IONFOCUS    :      90.157
 ENTRANCE_LE :      38.000
 EMVOLTS     :    1658.824
                               Actual EMV  :    1541.18
                               GAIN FACTOR :       1.02
 AMUGAIN     :    1702.000
 AMUOFFSET   :     121.188
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      16.063
 MASSGAIN    :    -725.000   
 MASSOFFSET  :     -41.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------