                           INSTRUMENT CONTROL PARAMETERS:    Maestro
                           -----------------------------------------

   C:\MSDCHEM\1\METHODS\Drug2012.M
      Tue Sep 02 09:57:43 2014

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

Oven
Equilibration Time                           4 min
Max Temperature                              425 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    50 C for 1 min
    then 10 C/min to 300 C for 14 min
Run Time                                     40 min




Front Injector
Syringe Size                                 10 ?L
Injection Volume                             1 ?L
Solvent A Washes (PreInj)                    10
Solvent A Washes (PostInj)                   12
Solvent A Volume                             8 ?L
Solvent B Washes (PreInj)                    0
Solvent B Washes (PostInj)                   0
Solvent B Volume                             8 ?L
Sample Washes                                3
Sample Wash Volume                           3 ?L
Sample Pumps                                 6
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 ?L/min
Solvent Wash Dispense Speed                  6000 ?L/min
Sample Wash Draw Speed                       300 ?L/min
Sample Wash Dispense Speed                   6000 ?L/min
Injection Dispense Speed                     300 ?L/min
Viscosity Delay                              0 sec
Sample Depth                                 Disabled
Injection Type                               Standard
L1 Airgap                                    0.2 ?L


Sample Overlap
Sample overlap is not enabled



Front SSZ Inlet He
Mode                                         Split
Heater                                       On    250 C
Pressure                                     On    7.6522 psi
Total Flow                                   On    41.952 mL/min
Septum Purge Flow                            On    0.95207 mL/min
Gas Saver                                    On    15 mL/min After 3 min
Split Ratio                                  40 :1
Split Flow                                   40 mL/min


Thermal Aux 1 (MSD Transfer Line)
Heater                                       On
Temperature Program                          On
    280 C for 0 min
Run Time                                     40 min


Column #1
Agilent 19091S-433: 1974.51236
HP-5MS  5% Phenyl Methyl Silox
325 C: 30 m x 250 ?m x 0.25 ?m
In: Front SSZ Inlet He
Out: Vacuum

(Initial)                                    50 C
Pressure                                     7.6522 psi
Flow                                         1 mL/min
Average Velocity                             36.445 cm/sec
Holdup Time                                  1.3719 min
Flow Program                                 On
    1 mL/min for 0 min
Run Time                                     40 min



Signals
Signal #1: Test Plot                         Save Off
                                             50 Hz
Signal #2: Test Plot                         Save Off
                                             50 Hz
Signal #3: Test Plot                         Save Off
                                             50 Hz
Signal #4: Test Plot                         Save Off
                                             50 Hz

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 4.00 min

EMV Mode                 : Gain Factor
Gain Factor              : 1.00
Resulting EM Voltage     : 1529

[Scan Parameters]

Low Mass                 : 35.0
High Mass                : 500.0
Threshold                : 150
Sample #                 : 2       A/D Samples    4

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US12051904
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      26.772
 IONFOCUS    :      90.157
 ENTRANCE_LE :      28.500
 EMVOLTS     :    1294.118
                               Actual EMV  :    1529.41
                               GAIN FACTOR :       0.97
 AMUGAIN     :    2225.000
 AMUOFFSET   :     120.313
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      18.573
 MASSGAIN    :    -446.000   
 MASSOFFSET  :     -38.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                              END OF INSTRUMENT CONTROL PARAMETERS
                              ------------------------------------