                          INSTRUMENT CONTROL PARAMETERS:    Maestro 2
                          -------------------------------------------

   C:\MSDCHEM\1\METHODS\SCRIN.M
      Fri Sep 04 18:21:07 2015

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

Oven
Equilibration Time                           0,2 min
Max Temperature                              325 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    150 C for 2 min
#1    then 25 C/min to 285 C for 0 min
#2    then 10 C/min to 300 C for 5 min
Run Time                                     13,9 min
5 min (Post Run)                             300 C
Cryo                                         Off

Front Injector
Syringe Size                                 10 ?L
Injection Volume                             2 ?L
Solvent A Washes (PreInj)                    2
Solvent A Washes (PostInj)                   5
Solvent A Volume                             8 ?L
Solvent B Washes (PreInj)                    2
Solvent B Washes (PostInj)                   5
Solvent B Volume                             8 ?L
Sample Washes                                1
Sample Wash Volume                           8 ?L
Sample Pumps                                 3
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 ?L/min
Solvent Wash Dispense Speed                  6000 ?L/min
Sample Wash Draw Speed                       300 ?L/min
Sample Wash Dispense Speed                   6000 ?L/min
Injection Dispense Speed                     300 ?L/min
Viscosity Delay                              0 sec
Sample Depth                                 Disabled
Injection Type                               Standard
L1 Airgap                                    0,2 ?L

Sample Overlap
Sample overlap is not enabled

Front SSZ Inlet He
Mode                                         Splitless
Heater                                       On    250 C
Pressure                                     On    16,008 psi
Total Flow                                   On    52,43 mL/min
Septum Purge Flow                            On    1,2299 mL/min
Gas Saver                                    On    20 mL/min After 2 min
Purge Flow to Split Vent                     50 mL/min at 1,5 min

Thermal Aux 1 (MSD Transfer Line)
Heater                                       On
Temperature Program                          On
    280 C for 0 min
Run Time                                     13,9 min

Column #1
Agilent 19091S-433: 2192.53941
HP-5MS  5% Phenyl Methyl Silox
325 C: 30 m x 250 ?m x 0,25 ?m
In: Front SSZ Inlet He
Out: Vacuum

(Initial)                                    150 C
Pressure                                     16,008 psi
Flow                                         1,2 mL/min
Average Velocity                             41,683 cm/sec
Holdup Time                                  1,1995 min
Flow Program                                 On
    1,2 mL/min for 0 min
Run Time                                     13,9 min
5 min (Post Run)                             2 mL/min

Signals
Signal #1: Test Plot                         Save Off
                                             50 Hz

Signal #2: Test Plot                         Save Off
                                             50 Hz

Signal #3: Test Plot                         Save Off
                                             50 Hz

Signal #4: Test Plot                         Save Off
                                             50 Hz

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 3.50 min

EMV Mode                 : Relative
Relative Voltage         : 200
Resulting EM Voltage     : 2129

[Scan Parameters]

Low Mass                 : 44.0
High Mass                : 550.0
Threshold                : 0
Sample #                 : 2       A/D Samples    4
Plot 2 low mass          : 50.0
Plot 2 high mass         : 550.0

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US12371902
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      25.432
 IONFOCUS    :      90.157
 ENTRANCE_LE :      32.000
 EMVOLTS     :    1929.412
                               Actual EMV  :    2129.41
                               GAIN FACTOR :       3.76
 AMUGAIN     :    1188.000
 AMUOFFSET   :     164.688
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      17.569
 MASSGAIN    :   -1083.000   
 MASSOFFSET  :     -35.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                              END OF INSTRUMENT CONTROL PARAMETERS
                              ------------------------------------