                                         INSTRUMENT CONTROL PARAMETERS:    GC-MS
                                         ---------------------------------------

   C:\MSDCHEM\1\METHODS\KATAEV-M1.M
      Mon Mar 14 15:17:44 2016

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  GC ALS
Mass Spectrometer :  Enabled


 No Sample Prep method has been assigned to this method.

Oven
Equilibration Time                           0.25 min
Max Temperature                              450 degrees C
Slow Fan                                     Disabled
Oven Program                                 On
    70 C for 2 min
    then 20 C/min to 280 C for 15.5 min
Run Time                                     28 min
Cryo                                         Off

Front Injector
Syringe Size                                 10 ?L
Injection Volume                             1 ?L
Solvent A Washes (PreInj)                    5
Solvent A Washes (PostInj)                   5
Solvent A Volume                             8 ?L
Solvent B Washes (PreInj)                    5
Solvent B Washes (PostInj)                   5
Solvent B Volume                             8 ?L
Sample Washes                                0
Sample Wash Volume                           8 ?L
Sample Pumps                                 5
Dwell Time (PreInj)                          0 min
Dwell Time (PostInj)                         0 min
Solvent Wash Draw Speed                      300 ?L/min
Solvent Wash Dispense Speed                  6000 ?L/min
Sample Wash Draw Speed                       300 ?L/min
Sample Wash Dispense Speed                   6000 ?L/min
Injection Dispense Speed                     300 ?L/min
Viscosity Delay                              0 sec
Sample Depth                                 -1 mm
Injection Type                               Standard
L1 Airgap                                    0.2 ?L

Sample Overlap
Sample overlap is not enabled

Front SS Inlet He
Mode                                         Split
Heater                                       On    250 C
Pressure                                     On    14.091 psi
Total Flow                                   On    19.5 mL/min
Septum Purge Flow                            On    3 mL/min
Gas Saver                                    Off
Split Ratio                                  10 :1
Split Flow                                   15 mL/min

Thermal Aux 2 (MSD Transfer Line)
Heater                                       On
Temperature Program                          On
    280 C for 0 min
Run Time                                     28 min

Column #1
Agilent 19091S-433HP-5MS  5% Phenyl Methyl Silox
325 C: 30 m x 250 ?m x 0.25 ?m
In: Front SS Inlet He
Out: Vacuum

(Initial)                                    70 C
Pressure                                     14.091 psi
Flow                                         1.5 mL/min
Average Velocity                             45.066 cm/sec
Holdup Time                                  1.1095 min
Flow Program                                 On
    1.5 mL/min for 0 min
Run Time                                     28 min

Valve 1
Name                                         ?
Gas Sampling Valve
GSV Loop Volume                              1 mL
Load Time                                    0.5 min
Inject Time                                  0.5 min

Signals
Signal #1: Test Plot                         Save Off
                                             50 Hz

Signal #2: Test Plot                         Save Off
                                             50 Hz

Signal #3: Test Plot                         Save Off
                                             50 Hz

Signal #4: Test Plot                         Save Off
                                             50 Hz

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 3.00 min

EMV Mode                 : Absolute
Resulting EM Voltage     : 1600

[Scan Parameters]

Low Mass                 : 42.0
High Mass                : 650.0
Threshold                : 50
Sample #                 : 2       A/D Samples    4
Plot 2 low mass          : 50.0
Plot 2 high mass         : 550.0

[MSZones]

MS Source                : 230 C   maximum 250 C
MS Quad                  : 150 C   maximum 200 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US12458805
                        -----------------------------

 Trace Ion Detection is OFF.

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      33.306
 IONFOCUS    :      90.157
 ENTRANCE_LE :      25.500
 EMVOLTS     :    1400.000
                               Actual EMV  :    1600
                               GAIN FACTOR :       1.88
 AMUGAIN     :    2326.000
 AMUOFFSET   :     119.875
 FILAMENT    :       1.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      19.827
 MASSGAIN    :    -457.000   
 MASSOFFSET  :     -41.000   

                           END OF TUNE PARAMETERS
                      ------------------------------------



                                           END OF INSTRUMENT CONTROL PARAMETERS
                                           ------------------------------------