                                                                       INSTRUMENT CONTROL PARAMETERS:    only_msd
                                                                       ------------------------------------------

   C:\MSDCHEM\1\METHODS\AD2013.M

Control Information
------- -----------
       Sample Inlet : Other/None
   Injection Source : Manual
 Injection Location : ALS
             Use MS : Yes

                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : atune.u
Acquistion Mode          : Scan
Run Length               : 20.00 min


MS Information
-- -----------

Solvent Delay            : 5.00 min

EM Absolute              : False
EM Offset                : 200
Resulting EM Voltage     : 1823.5

[Scan Parameters]

Low Mass                 : 40.0
High Mass                : 600.0
Threshold                : 150
Sample #                 : 2       A/D Samples    4
Plot 2 low mass          : 50.0
Plot 2 high mass         : 550.0

[MSZones]

MS Quad                  : 150 C   maximum 200 C
MS Source                : 230 C   maximum 250 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN:  
                        -----------------------------

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      34.814
 IONFOCUS    :      89.208
 ENTRANCE_LE :       9.500
 EMVOLTS     :    1623.529
 AMUGAIN     :    2576.000
 AMUOFFSET   :     130.000
 FILAMENT    :       1.000
 DCPOLARITY  :       1.000
 ENTLENSOFFS :      20.329
 MASSGAIN    :     698.000   
 MASSOFFSET  :      -8.000   


                           END OF TUNE PARAMETERS
                      ------------------------------------



                                                                          END OF INSTRUMENT CONTROL PARAMETERS
                                                                          ------------------------------------