                        INSTRUMENT CONTROL PARAMETERS:    Instrument #1
                        -----------------------------------------------

   C:\MSDCHEM\1\METHODS\NARCO_MET2_600.m

Control Information
------- -----------

Sample Inlet      :  GC
Injection Source  :  Manual
Injection Location:  Rear
Mass Spectrometer :  Enabled

=============================================================================
                                6850 GC METHOD
=============================================================================


OVEN
   Initial temp:  100 'C (On)              Maximum temp:  325 'C
   Initial time:  3.00 min                 Equilibration time:  0.00 min
   Ramps:
      #  Rate  Final temp  Final time
      1 25.00      300       20.00
      2   0.0(Off)
   Post temp:  0 'C
   Post time:  0.00 min
   Run time:  31.00 min


INLET (SPLIT/SPLITLESS)                 DETECTOR (NO DET)
   Mode:  Pulsed Splitless                 Temperature:  300 'C (On)
   Initial temp:  280 'C (On)
   Pressure:  10.49 psi (On)
   Pulse pressure:  20.0 psi
   Pulse time:  0.50 min
   Purge flow:  40.0 mL/min
   Purge time:  0.50 min
   Total flow:  43.6 mL/min
   Gas saver:  On
   Saver flow:  20.0 mL/min
   Saver time:  2.00 min
   Gas type:  Helium


COLUMN                                  SIGNAL
   Capillary Column                        Data rate:  0.1 Hz
   Model Number:  Agilent 19091S-433E      Type:  col comp
   HP-5MS  5% Phenyl Methyl Siloxane       Save Data:  Off
   Max temperature:  325 'C                Zero:  0.0 (Off)
   Nominal length:  30.0 m                 Range:  0
   Nominal diameter:  250.00 um            Fast Peaks:  Off
   Nominal film thickness:  0.25 um        Attenuation:  0
   Mode:  constant flow
   Initial flow:  1.0 mL/min
   Nominal init pressure:  10.50 psi
   Average velocity:  37 cm/sec
   Source:  Inlet
   Outlet:  MSD
   Outlet pressure:  vacuum


COLUMN COMP
   Use:  Valve Box Heater


VALVES                                  POST RUN
                                           Post Time: 0.00 min


TIME TABLE
   Time       Specifier                     Parameter & Setpoint




                               GC Injector


     Front Injector:
No parameters specified

     Back Injector:
No parameters specified

 Column 1 Inventory Number : AB001
                                MS ACQUISITION PARAMETERS


General Information
------- -----------

Tune File                : urtune.u
Acquistion Mode          : Scan


MS Information
-- -----------

Solvent Delay            : 4.00 min

EM Absolute              : False
EM Offset                : 0
Resulting EM Voltage     : 1623.5

[Scan Parameters]

Low Mass                 : 40.0
High Mass                : 600.0
Threshold                : 100
Sample #                 : 2       A/D Samples    4

[MSZones]

MS Quad                  : 150 C   maximum 200 C
MS Source                : 230 C   maximum 250 C


                             END OF MS ACQUISITION PARAMETERS


                              TUNE PARAMETERS for SN: US30945420
                        -----------------------------

 EMISSION    :      34.610
 ENERGY      :      69.922
 REPELLER    :      28.783
 IONFOCUS    :      90.157
 ENTRANCE_LE :      14.500
 EMVOLTS     :    1623.529
 AMUGAIN     :    2469.000
 AMUOFFSET   :     131.000
 FILAMENT    :       2.000
 DCPOLARITY  :       0.000
 ENTLENSOFFS :      19.075
 MASSGAIN    :     392.000   
 MASSOFFSET  :     -10.000   


                           END OF TUNE PARAMETERS
                      ------------------------------------



                              END OF INSTRUMENT CONTROL PARAMETERS
                              ------------------------------------